Maintenance

Custom Designed Office Solutions

  • Compressor Maintenance
  • Vacuum Maintenance
  • Amalgam Separator Service
  • Autoclaves

Compressor Maintenance:

Oil-Less Head(s)
  • Change Coalescing Filter and clean out coalescing
    filter bowl.
  • Check storage tank for water and drain where
    applicable.
  • Change air intake filters.
  • Check start and stop pressures. Adjust as needed.
  • Check incoming line voltage. Advise and make
    recommendations accordingly.
  • Perform overall inspection and cleaning of external
    components.
Oil-Head(s)
  • Change Coalescing Filter and clean out coalescing
    filter bowl.
  • Drain and flush head of oil and refill with fresh oil.
    Check storage tank for water and drain where
    applicable
  • Change air intake filters.
  • Check start and stop pressures. Adjust as needed.
  • Check incoming line voltage. Advise and make
    recommendations accordingly.
  • Perform overall inspection and cleaning of external
    components.

Vacuum Maintenance:

Water Ring (Wet)
  • Change solids collector.
  • Check foam air intake filter – Change as needed.
  • Check incoming water supply.
  • Change main water filter in Water Filter Solenoid
    where applicable.
  • Check vacuum level and settings. Adjust as needed.
  • Check for substantial vacuum leaks within the office if suspected.
  • Check incoming line voltage. Advise and make recommendations accordingly.
Dry Vacuum (Not Surgical Suction Units)
  • Check oil level and change oil in motor (Recommended
    once every ten years).
  • Perform visual inspection of drive belt and change as
    needed (usually once every 4 to 5 years)
  • Perform visual inspection of drive belt and change as
    needed (usually once every 4 to 5 years)
  • Inspect drain and drain check valve.
  • Check exhaust “P” trap for clear draining.
Surgical Suction Units
  • Change exhaust muffler(s).
  • Change inlet and outlet filters.
  • Check incoming line voltage. Advise and make recommendations accordingly.
  • Check start/stop pressures and pressure switch. Adjust as needed.
  • Check vanes in head(s) if an issue is suspected. Replace as needed.
  • Perform overall inspection and cleaning of external components.

Amalgam Separator Service:

  • Change amalgam separator and check for leaks.
  • Package up and prepare return cartridge for recycling when ordered through Dependable Dental.

Autoclaves:

  • Drain and wipe/clean unit.
  • Run 6 minute cleaning cycle.
  • Drain and refill and run 3 minute flush cycle.
  • Inspect door gasket and spacer, door springs, drain andfill filter screens. Replace as needed.
  • Clean and/or flush fill solenoid as needed.
  • Check trays and tray rack. Clean/wipe down as needed.